RHD Recording Headstages

We offer several types of RHD headstages that allow users to connect to electrodes (not included) and record electrophysiological signals from up to 128 low-noise amplifier channels.

Each headstage contains an Intan RHD amplifier chip, a few support components, a small connector to mate with a digital interface cable, and a connector to mate with a variety of commercial recording electrodes. Multiple headstages may be connected to an RHD recording controller or USB interface board using thin, flexible SPI interface cables. All signals are amplified and digitized directly on the headstage, so data on the interface cables is purely digital and immune to noise pickup.

Every RHD headstage has the following capabilities when used in an RHD recording system:

  • Monitor and record signals from all amplifier channels at sampling rates up to 30 kSamples/s per channel.
  • Reconfigure amplifier analog bandwidth from software over a range of 0.1 Hz to 20 kHz.
  • Measure in situ electrode impedances at user-selectable frequencies.

Note: RHD headstages do not work with the RHS stim/recording controller. The RHD and RHS systems are not compatible

RHD 128-Channel Recording Headstage

Our 128-channel recording headstage has two 64-pin Molex SlimStack electrode connectors that mate with the following silicon probes:

Explore more details on the operation of the 128-channel headstage.

RHD 128-channel headstage
Technical Specs
Part #C3316 RHD 128ch
headstage inputs
128 unipolar inputs, common reference, ground
electrode connector
2x Molex SlimStack 502426-6410
data cable connector
12-pin RHD SPI cable connector; high-speed serial digital data
size
35 mm x 21 mm
mass
2.05 grams
Headstage Details

The zero-ohm jumper R0 on the bottom side of the headstage may be removed to disconnect the reference electrode from ground. Two mounting holes are provided. Center-to-center spacing of the two Molex connectors is 8.8646 mm (0.349 inches).